Specifications
Capacity: 16 Sample Plates (96 or 384-well)
Capillary Length: 36 or 50 cm
For Use With (Equipment): Real-Time PCR Systems or Thermal Cyclers, Sequencing Data Analysis
Number of Capillaries: 96 Capillaries
Product Line: Applied Biosystems
Read Length: 400 to 900 Bases
Throughput: 96
Type: DNA Analyzer
Duration: 1 Year Limited Warranty on Parts and Labor
External Computer: Computer and Flat Panel Monitor
Format: 384-well plate, 96-well plate
High-throughput Compatibility: High-throughput Compatible
Memory16GB (2x8GB) 1600MHz DDR3 Non-ECC
Operating Environment: Humidity: 20%-80% (non-condensing), Temperature: 18°C-30°C, Humidity: 20 to 80% (non-condensing)
Operating System: Microsoft Windows 10
Optics: In-capillary detection by dual-side illumination, Backside-thinned CCD for maximum optical sensitivity, Multiple Filter Sets: Z, E, G5, G5-RCT, Any4Dyeset, Any5Dyeset and S (optional)
Performance: XLRseq: 50cm – 900 QV20 bases/3 hours, RapidSeq: 36cm – 550 QV20 bases/36 min, GeneMapper™ Software: 36cm – up to 500 bp res. w/ 0.15 bp sizing precision in 34 min, FastSeq: 50cm – 700 QV20 bases/1 hour, StdSeq: 36cm – 700 QV20 bases/1 hour, LongSeq: 50cm – 850 QV20 bases/2 hours
Plate Handling: Plate Seal: Automatic on-board piercing (septa or polypropylene heat seal), Internal bar code reader, Integrated auto sampler and plate stacker
PolymerPOP-7, POP Conformational Analysis Polymer
Polymer Consumption: 36 cm: 200μl⁄run, 50 cm: 250μl⁄run
Power Requirements: Voltage: 200-220 V or 230-240 V ±10%, Power Dissipated: About 2000 watts, Current: 15 amps max. Requires 30 amp receptacle w/in 3 m (10 ft) of instrument back.
Quantity1 analyzer
Reagent Handling: On-board reagents for up to 48 hrs unattended operation (with certain run modules), Automated polymer replenishment pump for increased throughput, Run modules: 5 sequencing, 1 fragment analysis and optional SNPlex™
Sample Volume: 10-50 μL for 96-well plate, 5-30 μL for 384-well plate
Shipping Condition: Room Temperature
Temperature Range (Metric): Temperature-controlled oven (18-70°C range) and detection cell heater
Dimensions: 100 cm (W) x 65 cm (D) x 93 cm (H)
Weight: 180 kg (computer – 30 kg)
Unit Size: 1 analyzer